SYNTHESIS Q.A.D. Rev. 02

 

 

“SYNTHESIS Q.A.D.” rev. 02 is a full programmable modular system for anterior segment. This unit allows the surgeon to perform the phacoemulsification, infusion/aspiration, vitrectomy for anterior or posterior (depending of the vitrectomy model). The unit automaticaly detect the probe insert or Diathermy (eso or endo) or  the capsulotomy handpieces; all the functions are individually selectable by panel switch icons or cyclical selectable by footswitch (Black switch toggle ). Every function is showed on the panel LCD display and audible by a sound message.

The Q.A.D. is a based system in a double intervention, electronic and hydraulic. The electronic  uses a microprocessor which with a proper dedicated software detects the pressure trends through the sensor in the aspiration line. A rapid transient in the vacuum line actives the software which gives proper signals in order to minimize the collapse of the anterior chamber. 

All parameters are fully programmable and storable in memory for 5 Surgeon.

According to the vacuum level the microprocessor gives the correct venting time. 

 “SYNTHESIS Q.A.D.” Phaco emulsification & Fragmentation 40 kHz (about) Phaco emulsification Standard (85gr) and Phaco fragmentation Light (35gr) both model are in titanium.

 

 

It is possible to apply on the handpiece different U/S phaco needles :

   

LOW TURBOLENCE PHACO NEEDLES

  • Ø 1.1mm  19 gauge

  • Available in 0°, 15°, 30° and 45°angles

  • Special needle (Miniflower) for high performance and better chamber stability

  • Designed to reduce turbulence and prevent bubbles  

 

SMALL INCISION PHACO NEEDLES  

  • For a better chamber stability

  • Ø 0,9mm 20 gauge

  • Available in 0°, 15°, 30° and 45° angles straight or curved.

  • Bubble free

  • Designed to give a significant control through a 2.5mm incision  

 

FRAGMENTATION PHACO NEEDLES  

  • Fragmentation Phaco Needle 20Ga 15°

  • Fragmentation Phaco Needle 20Ga 30°

  • Fragmentation Phaco Needle 20Ga 45°

 

The U/S handpieces are: Standard and Light (85gr and 45gr). A continuous auto-tuning 42 KHz and 100 micro stroke lengths maintains stability and ensures a constant needle elongation  during  the surgeon act.

   

“SYNTHESIS Q.A.D.” Infusion/Aspiration process removes the lens through the I/A instrument evacuation port. Aspiration is obtained through the flow rate control , thank to the rotary movement of the peristaltic roller bar pump on the silicone tube creates the vacuum and also provides predictable and stable aspiration. The vacuum in the aspiration line is from 4 to a  max. 500mm Hg . The maximum flow rate , produced by the peristaltic pump is  60 cc/min.

   

“SYNTHESIS Q.A.D.” The Synthesis Q.A.D. in vitrectomy mode allows the operator to perform anterior and posterior segment using the power handpiece with Guillotine tip. The  power pressure is obtained by a  compressor located on the unit. The cutting rate  is  from 50 beats per minute to a maximum of 1000 per minute.

It’s possible use different models for anterior and posterior segment(20psi).

   

“SYNTHESIS Q.A.D.” offers  RF  diathermy for a wide range of bipolar forceps, pencils, eraser or different endo-probe shape. 

The SYNTHESIS QAD Rev.02 can store the function’s parameters and the preferred manage mode for all the unit’s function up to 5 different surgeons; so 5 different surgeon can fully configure the unit for their use at the same time. When the surgeon recall his stored parameter he can modify them and use the unit with this new parameter. The unit will remember this new set while it’s switch on; if the surgeon store this new set of parameter, using the right button on the front panel, the unit will remember this set even if it’s turn off.

Technical Information

Electrical

LineVoltage:                                            100-120/200-230VAC,50÷60Hz

Maximum current                                    1 A / 0.5 A

Fuse (two Used)                                      F4A125VAC / F2A250VAC 5mmx20mm T listed  

Protection against electrical shock         Type/B equipment

Protection                                                Class II

Protection against ingress of water        Ordinary

Operation mode                                       Continuous

 

Diathermy

Type                                              Bipolar

Output Power                              7 Watt max

Output frequency                         550 kHz

   

Vacuum

Vacuum Range                             4 to 500 mmHg

 

Flow

Flow range                                    10 to 60 cc/min.        

 

Vitrectomy

Pneumatic handpiece

Work pressure                               20 psi

Cutting Range                                50 to 1000 cuts per minute     

 

Ultrasonic handpieces Standard or Ligth

Operating frequency                       38Khz to 42 kHz

Handpiece Type                              Piezokinetic

Sterilization mode                          Autoclavable                          

 

Environmental Limitations

Non operating

  •       Ambient temperature from          – 40°C to 60°C (- 40°F to 140°F)

  •       Relative humidity                        10% to 95%     Without condensation

  •       Altitude                                        12191mt  (40000 feet)

 

Operating  

  •       Ambient temperature from          10°C  to 40°C (50° F to 95° F)

  •       Relative humidity                         10% to 95% Without condensation

  •       Altitude                                         1828mt (6000 feet)

   

Dimensions

Height                                                           17 cm

Base                                                              31 cm

Depth                                                            34 cm

Weight                                                          12 Kg Unpacked

Weight                                                          20 Kg Packed

 

More Information

Complete unit for anterior segment 

 

·      Fully programmable up to 5 surgeons  

·      Voice message on mode changes  

·      Phacoemulsification:

·      Continuous mode

·      Linear control of U/S power

·      On/Off control of U/S power

·      Pulsed mode

·      Linear control of U/S power

·      On/Off control of U/S power

·      Burst Mode

·      On/Off control of U/S power

 

·      Irrigation/Aspiration:

·      Linear control of vacuum

·      On/Off control of vacuum

·      Linear control of flow rate

·      On/Off control of flow rate

·      Venturi mode only in phaco mode

·      Capsule polishing mode

 

·      Vitrectomy

·      Up to 1000 CPM

 

·      R.F. diathermy:

·      On Off control

·      Linear control

 

·      Test:

·      Vacuum test

·      Venting

·      U/S test

 

·      Capsulotomy (on request)

 

·      Automatic IV pole (on request)

 

·      Double linear contol footswitch

·      Function select mode

·      Continuous irrigation

·      Reflux

Technical specifications are for reference only and are subject to change without notice.