SYNTHESIS Q.A.D. Rev. 03

Cold Phaco- same energy, high vacuum,
no incision burns, stable AC
Surgicon Synthesis QAD Rev. 03 is
an ophthalmic surgical instrument manufactured to be durable,
reliable, safe and easy to operate. The Synthesis unit is designed
for use in the anterior segment that require lens fragmentation,
irrigation, aspiration as well as associated procedures such us
vitrectomy and coagulation. It has been developed to achieve mainly
two targets: to be user friendly and to allows the surgeon a great
control and flexibility; moreover this unit implements a innovative
cold phaco system.
Technical
Information
I/A Peristaltic system
Vacuum 0-500mmHg, 4mmHg increments
Flow
rate 10-60 ml/min, 1 ml/min increments
Reflux Selectable from dual linear control
pedal’s switch
Venting Selectable on the release of the dual
linear control pedal
Stability Automatic QAD system to for prevent
the collapse
Mode Linear Flow, Linear Vacuum, On/Off
Flow, On/Off Vacuum
Programmability Fully configurable up to 5 surgeons
Ultrasound Phaco
Frequency 42kHz piezo driver
Control Autotuning 35 times /sec, linear,
bilinear
Power 5% - 100% in steps of 5%
Mode Linear, On/Off, Pulsed linear, Pulsed
On/Off, and Burst mode
Pulse 1-15 PPS in steps of 1 PPS
Burst 100ms burst from 0.5Hz up to
continuous linearly controlled with bilinear pedal
Add. I/A
mode Venturi emulation
Cold
Phaco This Cold Phaco system allows the surgeon to
perform the surgeon act using bimanual
techniques with usual emulsification power
Cooling
rate 20-40 cr
Programmability Fully configurable up to 5 surgeons
Counter Effective time, 1sec increments
Diathermy mode
Frequency 1MHz
Power 7.5W on 100Ohm
Control Linear or On/Off
Programmability Fully configurable up to 5 surgeons
Vitrectomy mode
Air
source Internal compressor
Cutting
rate 50-1000cpm in steps 50cpm
Programmability Fully configurable up to 5 surgeons
Control On/Off
Capsule polishing
This function is specifically
design to polishing the surface of the posterior capsule allowing
the surgeon to touch the surface of the capsule and polish it.
Bilinear Pedal
Reflux.
Continuous
irrigation.
Mode
select.
Dual
linear movements.
RF Capsulotomy on request
Cart with I/V pole automatic
Power requirements
Line
Voltage 200-240VAC, 50..60Hz.
Consumption 0.5A.
Dimension and weight
Unit 400x400x200mm w/d/h 10kg.
Bilinear
pedal 240x280x210xmm w/d/h 2kg.
Standards
IEC601-1, EN60601-1, CE Mark.
Technical
specifications are for reference only and are subject to change
without notice |